Definition of Significant Platelet Clumping: Should We Review All Samples With a Platelet Clumping Flag From Automated Hematology Analyzer?
- Abstract
- Objectives: This study aimed to determine a definition for significant platelet clumping (PC) and evaluate the performance of the Sysmex XN instrument for detecting platelet clumps.
Methods: For part 1, 372 specimens with a 'PLT_clump?' flag in XN-9000 were classified into five groups according to the average number of PCs. We compared the initial platelet count (measured by XN-9000 using impedance method) and corrected platelet count (counted optically or re-analyzed by XN-9000 using vortexed or re-collected sample) of each group. For part 2, 1000 specimens with a PC flag divided into three subgroups {group N (PC = 0), Y (PC ≥ 1), and Z (microscopic fibrin clot)} and additional two groups {group S (PC(+) specimens without any flag and with flags of other categories) and group NC (negative control)} were collected. Positive predictive value (PPV), negative predictive value (NPV), sensitivity and specificity of PC detection of XN-9000 were obtained and the platelet counts and four indices (PDW, MPV, P_LCR, and PCT) of groups NC, N, Y, Z, and S were compared to detect PC more precisely.
Results: In part 1, all groups showed significant difference between the initial and corrected platelet counts. In part 2, PPV, NPV, prevalence, sensitivity, and specificity were 41.5%, 56.5%, 43.4%, 2.18%, and 98.3%, respectively. The platelet counts and four indices showed statistical differences for detecting PCs, and especially PDW and P_LCR were significantly smaller in group Z than group N or Y.
Conclusions: We suggest the definition of significant PC by the presence of at least three platelets. In addition, utilizing platelet-related indices should be developed to improve the efficiency of the PC detection.
- All Author(s)
- Suhyeon Woo
; Bohyun Kim
; Nam Hun Heo
; Min-Sun Kim
; Young Ahn Yoon
; Young-Jin Choi
- Intsitutional Author(s)
- 김보현; 윤영안; 최영진
- Issued Date
- 2024
- Type
- Article
- Keyword
- automated hematology analyzer; flag; platelet clumping; platelet count; pseudothrombocytopenia
- Publisher
- Blackwell Scientific Publications
- ISSN
- 1751-5521
; 1751-553X
- Citation Title
- International journal of laboratory hematology
- Citation Volume
- 47
- Citation Number
- 1
- Citation Start Page
- 79
- Citation End Page
- 86
- Language(ISO)
- eng
- DOI
- 10.1111/ijlh.14387
- URI
- http://schca-ir.schmc.ac.kr/handle/2022.oak/4745
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.