Objectives: This study aimed to determine a definition for significant platelet clumping (PC) and evaluate the performance of the Sysmex XN instrument for detecting platelet clumps.

Methods: For part 1, 372 specimens with a 'PLT_clump?' flag in XN-9000 were classified into five groups according to the average number of PCs. We compared the initial platelet count (measured by XN-9000 using impedance method) and corrected platelet count (counted optically or re-analyzed by XN-9000 using vortexed or re-collected sample) of each group. For part 2, 1000 specimens with a PC flag divided into three subgroups {group N (PC = 0), Y (PC ≥ 1), and Z (microscopic fibrin clot)} and additional two groups {group S (PC(+) specimens without any flag and with flags of other categories) and group NC (negative control)} were collected. Positive predictive value (PPV), negative predictive value (NPV), sensitivity and specificity of PC detection of XN-9000 were obtained and the platelet counts and four indices (PDW, MPV, P_LCR, and PCT) of groups NC, N, Y, Z, and S were compared to detect PC more precisely.

Results: In part 1, all groups showed significant difference between the initial and corrected platelet counts. In part 2, PPV, NPV, prevalence, sensitivity, and specificity were 41.5%, 56.5%, 43.4%, 2.18%, and 98.3%, respectively. The platelet counts and four indices showed statistical differences for detecting PCs, and especially PDW and P_LCR were significantly smaller in group Z than group N or Y.

Conclusions: We suggest the definition of significant PC by the presence of at least three platelets. In addition, utilizing platelet-related indices should be developed to improve the efficiency of the PC detection.
All Author(s)
Suhyeon Woo ; Bohyun Kim ; Nam Hun Heo ; Min-Sun Kim ; Young Ahn Yoon ; Young-Jin Choi
Intsitutional Author(s)
김보현; 윤영안; 최영진
Issued Date
2024
Type
Article
Keyword
automated hematology analyzer; flag; platelet clumping; platelet count; pseudothrombocytopenia
Publisher
Blackwell Scientific Publications
ISSN
1751-5521 ; 1751-553X
Citation Title
International journal of laboratory hematology
Citation Volume
47
Citation Number
1
Citation Start Page
79
Citation End Page
86
Language(ISO)
eng
DOI
10.1111/ijlh.14387
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/4745
Appears in Collections:
진단검사의학과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.