Standard ERG의 정상치 및 성별과 연령에 따른 변화(II) -Burian-Allen 각막전극에 의한 검사결과 및 ERG-jet 각막전극과의 비교-
- Alternative Title
- Normal Value of Standard Electroretinography and Change with Age and Sex(II)-Results Using Burian-Allen Electrode and Comparison with Results Using EFG-jet Electrode-
- Abstract
- Using Burian-Allen bipolar contact lens electrode, Authors examined the normal 120 eyes of 120 subjects(male 57, female 63), age ranged 7 to 83 years(mean 44.5 years), with the method which was same as the ISCEV standard. There was no statistically significant differences between the value obtained using Burian-Allen electrode and the value obtained using ERG-jet electrode(p>0.05). We expect that this data would be useful testing method to evaluate the retinal disease. And this data may provide new reference value in the study of electroretinogram.
- All Author(s)
- S. H. Choi
; Y. H. Ohn
; H. Shin
- Intsitutional Author(s)
- 온영훈
- Issued Date
- 1999
- Type
- Article
- Keyword
- Burian-Allen electrode; Electroretinogram
- Publisher
- 대한안과학회
the Korean Ophthalmological Society
- ISSN
- 0378-6471
; 2092-9374
- Citation Title
- 대한안과학회지
Journal of the Korean Ophthalmological Society
- Citation Volume
- 40
- Citation Number
- 1
- Citation Start Page
- 128
- Citation End Page
- 139
- Language(ISO)
- kor
- URI
- http://schca-ir.schmc.ac.kr/handle/2022.oak/4152
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.