SCHMC

Modified non-flared fully covered self-expandable metal stent for preoperative biliary drainage in pancreatic cancer

Metadata Downloads
Abstract
BACKGROUND AND AIM: Durable biliary drainage is essential in patients with pancreatic cancer in the era of effective neoadjuvant chemotherapy (NACT). Plastic stent (PS) tends to occlude easily, and current metal stents are designed for nonresectable cancer. We evaluated the efficacy of a modified non-flared fully covered self-expandable metal stent (M-FCSEMS) during the perioperative period in patients with resectable or borderline resectable pancreatic cancer. METHODS: Consecutive patients with resectable or borderline resectable pancreatic ductal adenocarcinoma had a 12-mm M-FCSEMS (M-FCSEMS group) or 7-Fr PS (PS group) placed for biliary decompression before NACT or curative-intent surgery. The primary outcome was the re-intervention rate, and secondary outcomes were technical success, stent patency, and adverse events (AEs). RESULTS: Endoscopic stent placement was technically successful in all 60 patients. Twenty-three patients underwent surgery and 37 NACT before surgery. Re-intervention was performed in 10.0% (3/30) of patients in the M-FCSEMS group and 36.7% (11/30) of patients in the PS group (P=0.030). The stent patency at 180days was 89.8% (95% confidence interval, 77.3-100.0%) in the M-FCSEMS group and 30.2% (95% confidence interval, 11.4-80.0%) in the PS group (P<0.0001). Stent-related AEs occurred in 10.0% (3/30) in the M-FCSEMS group and 40.0% (12/30) in the PS group (P=0.015). Surgery-related AEs occurred in 10.5% (2/19) and 14.3% (3/21) in the M-FCSEMS and PS groups (P=0.549). CONCLUSIONS: Modified non-flared fully covered self-expandable metal stents are effective and safe for durable biliary drainage in patients with resectable or borderline resectable pancreatic cancer during the perioperative period.
All Author(s)
I. S. Shin ; J. H. Moon ; Y. N. Lee ; J. C. Chung ; J. Jeong ; S. Y. Choi ; T. H. Lee ; J. K. Yang ; S. W. Cha ; Y. D. Cho ; S. H. Park
Intsitutional Author(s)
이태훈; 양재국
Issued Date
2022
Type
Article
Keyword
Covered self-expandable metal stent; Drainage; Pancreatic neoplasms
Publisher
Asian Pacific Association of Gastroenterology
Asian Pacific Association for the Study of the Liver
ISSN
0815-9319
Citation Title
Journal of Gastroenterology and Hepatology
Citation Volume
37
Citation Number
1
Citation Start Page
225
Citation End Page
232
Language(ISO)
eng
DOI
10.1111/jgh.15733